A-6 Bovie Medical Corporation
Guidance and manufacturer’ s declaration – electromagnetic immunity
The Aaron 950™ is intended for use in the electromagnetic environment listed below. The customer or
the user of the Aaron 950™ should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power sup-
ply input lines
IEC 61000-4-11
<5 %
U
t
(<95 % dip in
U
t
)
for 0.5 cycle
<40 %
U
t
(<60 % dip in
U
t
)
for 5 cycles
70 %
U
t
(<30 % dip in
U
t
)
for 25 cycles
<5 %
U
t
(>95 % dip in
U
t
)
for 5 sec
<5 %
U
t
(<95 % dip in
U
t
)
for 0.5 cycle
<40 %
U
t
(<60 % dip in
U
t
)
for 5 cycles
70 %
U
t
(<30 % dip in
U
t
)
for 25 cycles
<5 %
U
t
(>95 % dip in
U
t
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Aaron 950™ requires
continued operation during power
mains interruptions, it is
recommended that the Aaron
950™ be powered from an
uninterruptible power supply or a
battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE
U
t
is the a.c. mains voltage prior to application of the test level.