83
CALCulate Subsystem:
CALCulate:FUNCtion NULL
TOLerance
LIMit
AVERage
NONE
Description: Enable or disable the function of
the calculation.
Parameter:
NULL – enable null (relative) function;
TOL – enable tolerance function;
LIM – enable limit function;
AVER – enable average (static recording)
function;
NONE – disable all calculation and select the
internal trigger source.
Example: CALC:FUNC AVER
Note: The trigger source is changed to
immediate when the AVERage
function is enabled.
User can combine TOL/NULL,
LIM/NULL, AVER/NULL or
NULL/AVER functions, but the other
combinations are void.