EasyManua.ls Logo

B+K precision 830C - SCPI Commands

Default Icon
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
83
CALCulate Subsystem:
CALCulate:FUNCtion NULL
TOLerance
LIMit
AVERage
NONE
Description: Enable or disable the function of
the calculation.
Parameter:
NULL enable null (relative) function;
TOL enable tolerance function;
LIM enable limit function;
AVER enable average (static recording)
function;
NONE disable all calculation and select the
internal trigger source.
Example: CALC:FUNC AVER
Note: The trigger source is changed to
immediate when the AVERage
function is enabled.
User can combine TOL/NULL,
LIM/NULL, AVER/NULL or
NULL/AVER functions, but the other
combinations are void.