Bradley 233 - DC Voltage Test

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DC
level
:
to
be
between
12.
5
and
14.
5V.
Ripple
:
to
be
less
than
15mV
p-p.
4.
'-6V
Supply'
Measure
the
voltage
on
the
eyelet
near
R129
with
respect
to
the
nearest
end
of
R18
(earth).
DC
level
:
to
be
between
5.
7
and
6.7V.
Ripple
:
to
be
less
than
2.
5mV
p-p.
5.
'-18V
Supply'
Measure
the
voltage
on
the
eyelet
to
near
RH
corner;
(a)
w.r.t.
-12V
(adjacent
eyelet)
set
to
6V
using
RV443
(b)
w.r.t.
chassis :
ripple
20mV
p-p.
6.
Reset
Variac
to
240V.
5.5
DC
VOLTAGE
TEST
(Test
Equipment
3,
9,
10)
5.5.1
Output
Amplitudes
Select
level
:
max
-ve
+ve
slope
Ext
Trigger
Outputs:
Square-wave,
normal,
-ve.
Offset
zero,
Z
:
50p.
o
Measure
the
maximum
amplitude
of
each
output.
Using
RV443
adjust
the
value
of
the
smaller
output
to
be
between
-10.18
and
-10.
22V.
Record
the
maximum
and
minimum
voltages
of
each
range
on
both
outputs.
Check
that;
Minimum
amplitudes
are
between
0.1
and
0.
5V.
There
is
overlap
between
adjacent
ranges.
Maximum
amplitude
is
less
than
10.7V.
5.
5.
2
Output
Functions
Connect
output
1
to
DVM
(3).
Set
amp
to
maximum,
select
Comp,
(complementary)
checking
that
output
goes
to
zero
(±10mV).
Select
+ve
pulse.
Adjust
RV276
(output
1),
RV376
(output
2)
so
that
output
is
zero
(±20mV).
Select
'Norm'
and
record
the
+ve
amplitude
checking
that
its
value
is
the
same
but
of
oppo
site
sign
to
-ve
amplitude
(±0.
IV).
-
22