Guidance and manufacturer’s declaration – electromagnetic emissions
The NTF 3000 equipment is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the NTF 3000 should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment – guidance
RF Emissions
CISPR 11
Group 1
The ME equipment uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B Complies
Harmonic emissions
IEC 61000-3-2
Not
Applicable
The ME equipment is
solely battery powered.
Voltage uctuations/
icker emissions
Not
Applicable
Non-Life Support Equipment Separation Distance Calculation
(3Vrms / 3V/m compliance)
Rated maximum output
power of transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
in ISM bands
d = [
3,5
] P
V
1
80 MHz to
800 MHz
d = [
3,5
] P
E
1
800 MHz to 2.5 GHz
d = [
7
] P
E
1
0.01
0.12 0.12 0.23
0.1
0.37 0.37 0.74
1
1.17 1.17 2.33
10
3.69 3.69 7.38
100
11.67 11.67 23.33
Guidance and manufacturer’s declaration – electromagnetic immunity
The NTF 3000 is intended for use in the electromagnetic environment speci ed below.
The customer or the user of the NTF 3000 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
– guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete, or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Radiated RF
IEC 61000-4-3
3 V/m 80MHz to
2.5GHz
Complies
Field strengths outside the
shielded location from xed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than 3 V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance calculation
provided below. If a known
transmitter is present the
speci c distance can be
calculated using the equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not Applicable
(no electrical
cabling)
Electrical fast transient
IEC 61000-4-4
±2kV power line
±1kV I/O lines
Not Applicable
The ME equipment
is solely battery powered.
Surge
IEC 61000-4-5
±1kV
di erential
±2kV common
Not Applicable
Power frequency
magnetic eld
IEC 61000-4-8
3 A/m Complies
Power frequency
magnetic elds should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power
supply input lines
IEC 61000-4-11
>95% dip
0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Not Applicable
The ME equipment
is solely battery
powered.
English