10 11
Blank display. Thermometer
does not have power.
Please check if the
batteries have been
loaded correctly. Also
check polarity (<+> and
<–>) of batteries.
Contact customer
service if thermometer
still does not function.
System Error - self-check display
flashes continuously and will not
be followed by the ready beep
and the ready symbol.
If error persists.
If error still persists.
Wait 1 minute until the
thermometer turns off
automatically, then turn
on again
.
...rest the thermometer
by removing the batteries
and putting them back in.
...please contact
customer service.
Care and cleaning
Use an alcohol swab or cotton swab moistened with alcohol (70% isopropyl) to
clean the thermometer casing and the measuring probe. Ensure that no liquid
enters the interior of the thermometer. NEVER use abrasive cleaning agents,
thinners or benzene for cleaning and NEVER immerse the instrument in water
or other cleaning liquids. Wait 10 minutes after cleaning before taking a
temperature measurement.
NEVER insert a sharp object into the scanner area or any other open surface on
the thermometer.
Replacing the batteries
Replace with 2 new AAA batteries when the flashing battery symbol appears on
the LCD screen. To change the batteries, slide open the battery cover and
remove batteries. Replace the batteries being sure to align properly as indicated
inside the battery compartment. Remove the batteries from the product if not
used for extended periods of time in order to avoid damage to the thermometer
resulting from a leaking battery.
To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations.
Do not mix alkaline, standard, or rechargeable batteries
1 Year Limited warranty
This product comes with a 1 year limited warranty commencing on the date of
purchase. Within the warranty period we will eliminate, free of charge, any
defects in the appliance resulting from faults in materials or workmanship, by
replacing the complete appliance.
This warranty is applicable only for the appliance supplied by the appointed
distributor. This warranty does not cover: damage due to improper use, normal
wear or use as well as defects that have a negligible effect on the value or
operation of the appliance. The warranty becomes void if repairs are
undertaken by unauthorized persons and if original Braun parts are not used.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the use instructions, periodic re-adjustment
is not required. If at any time you question the accuracy, please contact
Consumer Relations.
Product specification
Type: Non-contact forehead thermometer
(BNT100)
Measuring range: Body: 34.0 °C – 43.0 °C (93.2 °F – 109.4 °F)
Surface: 0 °C – 100.0 °C
(32.0 °F – 212.0 °F)
Measuring location: Forehead
Resolution: 0.1 °C (0.1 °F)
Laboratory accuracy
- Body temperature mode: ±0.2 °C for the range 35.0 °C – 42.0 °C
(±0.4 °F for 95.0 °F – 107.6 °F)
±0.3 °C (±0.5 °F) for the range outside
35.0 °C – 42.0 °C (95.0 °F – 107.6 °F)
- Food/bath temperature mode: ±2.0 °C for the range 0.0 °C – 100.0 °C
(non-medical use) (±3.6 °F for 32.0 °F – 212.0 °F)
Clinical accuracy: Clinical performance and protocols are
available upon request
This thermometer displays a calculated
oral equivalent estimate
Display: Liquid Crystal Display, 4 digits plus
special icons
Acoustic: Audio:
Normal temp range = Green temp
range: 1 long beep for 0.4 second
duration
Fever = Yellow temp range: 1 long
beep for 0.4 second duration
High fever = Red: 10 short beeps
(0.02 seconds on, 0.13 seconds off)
Operating temperature: 15 °C – 40 °C (59.0 °F to 104 °F)
Automatic Switch-off: Within approx. 20 seconds
Weight: 93.2g (w/ batteries) and
71.5 (w/o batteries)
Size: 55mm x 38mm x 149mm
Long term storage ranges
Storage/transport temperature: -25 °C to 55 °C (-13 °F to 131 °F )
English