S
Sean JacobsSep 12, 2025
What to do if my Breas Clearway 2 Medical Equipment SD card is not recognized?
- RrpriceSep 12, 2025
Try using another SD card. If the issue persists, the Comms PCB or the Display PCB may be faulty and need to be replaced.
What to do if my Breas Clearway 2 Medical Equipment SD card is not recognized?
Try using another SD card. If the issue persists, the Comms PCB or the Display PCB may be faulty and need to be replaced.
What to do if my Breas Clearway 2 Medical Equipment SpO2 Sensor is not recognized?
Try using another sensor and/or another cable. If the issue continues, the Comms PCB or the Power PCB may be faulty and need to be replaced.
Display | LCD |
---|---|
Input Power | 100-240 VAC, 50/60 Hz |
Inspiratory Pressure Range | 4-30 cmH2O |
Expiratory Pressure Range | 4-20 cmH2O |
Inspiratory Time Range | 0.3-3.0 seconds |
Noise Level | < 30 dBA at 10 cmH2O |
Operating Temperature | +5°C to +40°C |
Humidity Range | 15-95% (non-condensing) |