10 Safety Information
iSleep 20 clinician’s manual
Doc. 003923 En-Us V-2
2.4 Usage of Patient Circuit
Only use the iSleep 20 with a mask and a patient tube recommended by
Breas Medical AB and your health care professional.
Do not breathe in the connected patient circuit unless the iSleep 20 is
turned on and operating properly.
Do not use patient hoses or tubes made of electrically conductive or static
material.
Always use a new mask and a patient tube recommended by Breas Medical
AB when the iSleep 20 is to be used by a new patient.
Patient connected parts and filter must be replaced regularly to ensure cor-
rect function of the iSleep 20. All replaced parts must be disposed of
according to local environmental regulations regarding the disposal of used
equipment and parts.
If the patient is using a full face mask (covering mouth and nose), the mask
must be equipped with a safety entrainment valve.
Make sure that the leakage ports are never blocked or obstructed. These
ports are used for ventilating the mask in order to prevent re-breathing of
exhaled air. Re-breathing of exhaled gases for longer than a few minutes
can, in some circumstances, lead to suffocation.
At low CPAP pressures, the air flow through the leakage ports may be inad-
equate to clear all exhaled gases from the mask. Some re-breathing may
occur.