• CPAP (Continuous Positive Airway Pressure)
Compatible Patient Circuits
The Vivo 45 should be used with a leakage circuit and a suitable patient interface or an
MPV circuit. See 9 Accessories and Parts, page 160 for detailed information about
approved patient circuits.
The patient circuit shall comply to ISO 17510. The leakage should be at least 12 l/min
at 4 cmH
2
O, to prevent rebreathing of exhaled air. The recommended leakage is 20 to
50 l/min at 10 cmH
2
O pressure.
Compatible Patient Interfaces
For invasive use, the patient interface may be a tracheostomy tube (cuffed or uncuffed).
For non-invasive use it may be a mask, mouthpiece or pillow interface. See the patient
interface’s instructions for use when selecting the interface to use.
Data Log
The Vivo 45 has an internal memory with a data log that holds the following data:
• Running hours
• Technical alarms
• Settings
• Asset data
• Treatment hours
• Treatment settings
• Device serial number
• Physiological alarms
• Detailed log, containing at least 24 h data of clinical data (monitored values)
• Breath log, containing at least 30 day data of (monitored values)
• Usage log (containing at least 1 year data of non-clinical events, alarms and settings)
The internal memory data is maintained also during power failure. The data can be
transferred to a computer, printed out, and analysed via Breas software products.
For more information about Breas software products, please contact your Breas
representative.
Expected Service Life
The expected service life of the Vivo 45 is 8 years.