Questions and Answers

  • M
    Michael LawsonAug 19, 2025
    Why won't my Brecknell 67 U turn on?
    • B
      brooke18Aug 19, 2025
      The issue might stem from several causes: the power cord could be improperly connected or not plugged in at all, the power outlet might not be providing electricity, or the battery could be discharged. To resolve this: * Check the power cord connections to ensure they are secure. * Verify that the power cord is plugged into a functioning power outlet. * Check the power source. * Replace the batteries if they are discharged.
  • D
    Diane HigginsAug 21, 2025
    What to do if my Brecknell 67 U Scales display shows '----'?
    • H
      hector50Aug 21, 2025
      This display indicates that the weight reading either exceeds the overload limit or the weight value, in the current unit of measure, exceeds 6 digits. To fix this, reduce the load on the scale until the weight value can be displayed or select a more appropriate unit of measure. You may also need to reset some parameters of CONFG or USER.
  • S
    Sherri LozanoAug 23, 2025
    How to fix 'Ad---' error on Brecknell Scales?
    • J
      James SimsAug 23, 2025
      The 'Ad---' error indicates that the loadcell wires connected to the indicator are incorrectly connected, shorted, or opened, or that the ADC or loadcell(s) are damaged. Ensure the wires are correctly connected and in good condition. Replacing the loadcell or ADC chip may be required.
  • E
    Elizabeth WallaceAug 26, 2025
    What does 'CAL.Er' mean on my Brecknell 67 U Scales and how to fix it?
    • C
      Cheryl StanleyAug 26, 2025
      A 'CAL.Er' error indicates a calibration problem. This could be due to incorrect input data, or the loaded weight being too small, too big, unstable, or non-linear. Ensure you input the correct data and load the correct weight onto the platform.
  • R
    Richard RichAug 27, 2025
    What to do if my Brecknell 67 U Scales cannot zero?
    • T
      Travis JacobsAug 28, 2025
      If your scale won't zero, it might be because the load on the scale exceeds the allowable limits, the load isn't stable, or there's load cell damage. Try removing the load from the scale and wait for the reading to stabilize.
  • D
    Dawn StrongAug 29, 2025
    What does 'EEP.E1' mean on my Brecknell 67 U Scales and how to solve it?
    • K
      khudsonAug 29, 2025
      The 'EEP.E1' error means that the CONFG or CAL parameters are not correctly set. To resolve this, reset the items in CONFG and perform a calibration.
  • K
    Kyle LindseySep 1, 2025
    What does 'EEP.E2' mean on my Brecknell 67 U Scales and how can I fix it?
    • M
      martinhectorSep 1, 2025
      The 'EEP.E2' error indicates that a USER parameter is not correctly set. To fix this, reset the items in USER.
  • C
    courtney90Sep 3, 2025
    What does 'CAP.ER' mean on Brecknell Scales?
    • C
      Carolyn BarnesSep 3, 2025
      The 'CAP.ER' error means that the capacity parameters are not correct. To fix it, set PRIM.N/PRIM.D/SECND.n to the correct number, ensuring that the capacity does not exceed 6 digits.
  • T
    Travis GravesSep 5, 2025
    Why do my Brecknell 67 U display '0----'?
    • D
      Derrick GibsonSep 5, 2025
      This indicates that the weight reading is below the Power On Zero limit. Install the platform on the scale and perform a zero calibration.
  • H
    Heather EvansSep 7, 2025
    Why do my Brecknell Scales display '____'?
    • K
      Kristina PrestonSep 7, 2025
      This display means the weight reading is below the Under load limit. Install the platform on the scale and perform a zero calibration.

Brecknell 67 U Series Specifications

General IconGeneral
MaterialStainless steel
Tare FunctionYes
Stabilization Time2 seconds
Display Digits5
DisplayLCD
Units of Measurementlb, kg
Auto-OffYes
Tare RangeFull Capacity