EasyManua.ls Logo

BRUEL & KJAER 2203 - Vibration Measurement Procedure

BRUEL & KJAER 2203
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.5.3.
Calibration
using
the
Internal
Reference
Voltage
The
stable
internal
voltage
0,2
dB
at
about
1000Hz)
in
the
2203
may
also
be used
for
calibration,
as
follows:
1.
Connect
the
accelerometer
to
the
2203
using
one
of
the
methods
shown
in
Fig.3.3.
If
an
integrator
is used, set
it
to
the
required
position
("Ace.",
"Vel.",
or
"Disp.").
2.
Pull
the
METER SWITCH
out,
and
set
it
to
"Batt."
to
check
the
batteries;
then
set
it
to
"Fast"
and
"Lin
.
".
(The
instrument
warm-up
time
is
20
s.)
3.
Set
the
ATIENUATORS
of
the
2203
so
that
"Ref
." is
inside
the
red
lines
(adjacent
to
the
black dot)
and
turn
the
ADJ.
potentiometer
with
a
small
screwdriver
to
give
one
of
the
following
readings,
depending
on
the
method
of
connection:
K-factor
of
"0"
dB
K-factor
of
"+
0,8"
dB
K-factor
of"+
1
,2"
dB
for
Integrator
ZR
0020
for
JJ
2614
for
JJ
2615
4.
The
2203
is
now
calibrated
so
that
a
reading
of
"94"
dB
corresponds
to
1 g
RMS
us-
ing
an
accelerometer
with
a
sensitivity
of
50
mV
/g.
For
accelerometers
of
different
sensitivities,
the
results
should
be
corrected
afterwards,
as
explained
in
section
3.7.3.
3.6.
VIBRATION
MEASUREMENTS
3.6.1.
General
When
using
the
2203
as a
preamplifier
and
measuring
amplifier
for
an
accelerometer,
it
should
be
remembered
that
the
accelerometer
has
a
high-frequency
resonance
(given
on
its
calibration
chart)
which
probably
lies
within
the
frequency
range
of
the
meter.
Mea-
surements
should
therefore
be
made
with
the
use
of
an
appropriate
supplementary
low-
pass
filter,
or
with
a
band-pass
filter
set
(see
Chapter
1
),
to
prevent
measuring
the
re-
sults
of
the
resonance,
where
it
is
thought
possible
that
the
signal
may
contain
such
high
frequencies.
The
lower
limiting
frequency
of
the
measuring
system
is
set
at
10Hz
by
the
2203.
For
detailed
instructions
on
the
use
of
the
accelerometer,
see its
Instruction
Manual.
The
2203
itself
should
be kept
out
of
the
vibration
environment.
3.6.2.
Procedure
for
Vibration
Measurements
20
1.
Calibrate
the
2203
(including
checking
the
batteries)
as
described
in
section
3.5,
and
mount
the
accelerometer
by
one
of
the
methods
described
in
its
Instruction
Man-
ual.
2.
Set
the
METER SWITCH
to
"Lin."
"Fast".
3.
Set
the
OUTPUT
ATIENUATOR
(transparent)
fully
clockwise
so
that
the
red
lines
are
adjacent
to
the
black spot.
4.
Adjust
the
INPUT
ATIENUATOR
(black)
to
give
the
highest
possible
meter
reading
without
over-deflecting
.

Other manuals for BRUEL & KJAER 2203

Related product manuals