14
CREST Mode (model 255 & 257 only)
Accuracy: Specified accuracy plus 150 digits for
changes > 5ms in duration
RECORD Mode (model 255 & 257 only)
Accuracy: Specified accuracy plus 100 digits
for changes > 100ms in duration
Audible Continuity Tester
Audible Threshold: Between 10 and 80
Response time: 32ms
Capacitance (model 252, 255 & 257 only)
Accuracies with film capacitor or better
Diode Tester
Test Current: 0.56mA typically
Open Circuit Voltage: < 1.8VDC typically
DC Current
RANGE Accuracy
A
0.5%+3d
60.00mA
0.5%+5d
1.7 mV/mA
600.0mA
0.5%+3d
1)
8A continuous, >8A to 15A for 30 sec. max
with 5 minutes cool down interval
AC Current
RANGE Accuracy
1)
8A continuous, >8A to 15A for 30 sec. max
with 5 minutes cool down interval
Temperature (model 252, 255 & 257 only)
K type thermocouple range & accuracy not
included
Logic Level Hz (mV Function)
RANGE
Accuracy: 0.03%+2d
Hz (Line) @ ACV, DCV, Current &
AutoCheck
TM
Function
Accuracy: 0.03%+3d
Non-Contact EF-Detection
Typical Voltage
Indication: Bar-graph segments & audible beep
tones proportional to the field strength
Detection Frequency: 50/60Hz
Detection Antenna: Top-right end of the meter
Probe-Contact EF-Detection: For more precise
indication of live wires, such as distinguishing
between live and ground connections, use the
Red (+) test probe for direct contact
measurements