26
Ohms
RANGE
Accuracy
1)
600.00
0.085%+10d
6.0000k
, 60.000k
0.085%+4d
600.00k
0.15%+4d
6.0000M
2)
1.5%+5d
60.000M
3) 4)
2.0%+5d
99.99nS
5)
1.0%+10d
Open Circuit Voltage: < 1.3VDC (< 1.5VDC for 600
range)
1)
Temperature Coefficient: 0.20 x (specified accuracy)/
C @ (-20
C ~ 18
C or 28
C ~
55
C)
2)
Constant Test Current: 0.1
µ
A Typical
3)
Constant Test Current: 0.01
µ
A Typical
4)
Specified accuracy adds 0.5% @ >50M
5)
Specified accuracy adds 30d @ <10nS
BeepLit
TM
Continuity Tester
Audible threshold: between 100
and 420
Response time < 100
s
Audible Indication: Beep Sound
Visible Response: LCD Backlight
BeepLit
TM
Diode Tester
RANGE
Accuracy
Test Current (Typical)
Open Circuit Voltage
3.0000V
1%+20d
0.35mA
< 3.1 VDC
Short-Beep-Alert Threshold: Drop Across 0.850V
BeepLitcontinuous ON Threshold: < 0.100V
Audible Indication: Beep Sound
Visible Indication: LCD Backlight
Capacitance
RANGE
Accuracy
1) 2)
10.00nF
1.0% + 10d
100.0nF~1000nF
1.0% + 2d
10.00
F~1.000mF
1.8% + 4d
10.00mF
2.0% + 4d
1)
Accuracies with film capacitor or better
2)
Temperature Coefficient: 0.20 x (specified accuracy)/
C @ (-20
C ~ 18
C or 28
C ~