1.8mm/s r.m.s. in the range 10 to1000 Hz
A-weighted sound
pressure level measured at 1meter
Table.7 Vibration and Noise data
2.8 Connection
Cooling water connection:
Inlet
Outlet
3/8inch
3/8inch
Vacuum connection:
Inlet
Outlet
BSJ600LC
ISO160
ISO100
BSJ1200LC
ISO250
ISO100
Table.8 Connection
3. Installation
3.1 Safety
WARNING
Obey the safety instructions in this s
ection and take note of suitable
precautions. Failure to observe these instructions may result in injury to
people and damage to equipment.
• The pump must be installed by suitably trained and instructed technician.
• Ensure that the installer is familiar with the safety procedures associated with the
product being pumped. Wear protective clothing when in contact with contaminated
parts. Disassemble and contaminated parts in the fume hood.
• Degassing and purging the pumping system before starting installation work.
• Before starting work, check that the required parts are ready and the type is correct.
• Make sure that no debris enters pump when installing.
• Other electrical components in the pumping system should be disconnected from the
power supply to prevent accidental activation.
• Ensure that the cooling air flow through the pump motor cooling fan cannot be
obstructed.
• All pump servicing points and oil level sight glasses should have suitable access.
• Ensure that all electrical cables and cooling water pipelines are in a safe position and the