Page 8
BUCK BioAire™ Pump Model B520, (APB-706000) complete
with standard 120VAC adapter/charger, instruction manual, and
APB-706100 Calibration Rotameter Assembly
BUCK BioAire™ Pump Model B520, (APB-706020) complete
with standard 230VAC adapter/charger, instruction manual, and
APB-706100 Calibration Rotameter Assembly
Buck Standard Charger, (APB-603900) supplied, 120VAC
Designed to charge the BioAire™ Pump Battery Pack (6 hours
charge time)
Buck Standard Charger, (APB-603910) supplied, 230VAC
Designed to charge the BioAire™ Pump Battery Pack (6 hours
charge time)
BUCK FastOne™ Charger (APB-601900) optional, 90-240VAC
Single Station Automatic One Hour Charger
BUCK BioAire™ Rechargeable Battery Pack 4.8VDC
(APB-108014) The pack of four cells can easily be changed by
opening the case and lifting the pack off the Velcro strip,
unplugging the connector and installing the new pack
Vinyl Tubing (APB-109132) A ten foot section of 1/2 inch ID vinyl
tubing for remote sample collection with Buck BioAire™ and
Air-O-Cell™ cassettes
Calibration Rotameter Assembly (APB-706100) fits on top of
Bioaerosol cassette for convenient flowrate setting and pump
calibration
Tripod for BioAire™ support (APB-104102) supports BioAire™ for
breathing zone and other elevated samples
*Air-O-Cell is a trademark of Zefon International
BioAire and FastOne are trademarks of A.P. Buck, Inc.
Parts and Accessories
Page 5
Audible Alert ON/OFF
The arrow keys will turn off the audible “beep” in areas where sound
is not appropriate. With the pump power ON, press the Down Arrow
Key for 4 (four) seconds or until two beeps acknowledge the disabling
of the beeper. Press the Up Arrow Key for 4 (four) seconds or until
three beeps signify the audible beeper is ON.
Low Battery Warning
When sampling pressing the down arrow key will light up the LED’s,
located on the right side of the unit, based on the battery charge. If all 5
LED light the batteries are 100%. If 4 light is 80%, 3 light = 60%, etc.
The low battery LED will begin flashing when there is only enough
power to collect a 10 minute sample.
Sample Timing Lock-Out
Once the Red low battery warning LED stops blinking and becomes
Solid Red, none of the timing keys will function and a new sampling
cannot be started. At this time the pump must be recharged or plugged
into the charger to continue sampling.
Tripod Mount
A connector located on the bottom of the unit provides easy
connection to most standard tripods.
Specifications
• Model: BioAire™ Model B520
• Flow Range 5 to 20LPM
• Accuracy ± 5% at Set Point
• Run Time 5 hours @ 15LPM
• Size 4”H x 6”W x 5.25D
(10.2 cm
H x 15.2 cm W x 13.3 cm D)
• Weight 22oz (624 g)
• Battery Voltage 4.8VDC
(120 or 230VAC standard charger included)
• Compatibility: Buck BioAire™ or Air-O-Cell™ cassettes
• Recharge Time 6 hours (1 hour with optional FastOne™
90-240VAC Charger)
• Marks