EasyManua.ls Logo

Buhler PKE 511 - User Manual

Buhler PKE 511
46 pages
Go to English
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Bedienungs- und Installationsanleitung
Installation- and Operation Instruction
Messgaskühler / Sample Gas Coolers PKE-5xx
BX440010 08/2013 Art. Nr. 90 31 105
Bühler Technologies GmbH, Harkortstr. 29, D-40880 Ratingen
Tel. +49 2102 49 89-0, Fax. +49 2102 49 89-20
Internet: www.buehler-technologies.com
Email: analyse@buehler-technologies.com
1
Lesen Sie die Bedienungsanleitung vor dem Gebrauch des Gerätes gründlich durch. Beachten Sie
insbesondere die Hinweise unter Gliederungspunkt 2. Andernfalls könnten Gesundheits- oder
Sachschäden auftreten. Die Bühler Technologies GmbH haftet nicht bei eigenmächtigen Änderungen
des Gerätes oder für unsachgemäßen Gebrauch.
Read this instruction carefully prior to installation and/or use. Pay attention particularly to all advises
and safety instructions to prevent injuries. Bühler Technologies can not be held responsible for
misusing the product or unreliable function due to unauthorised modifications

Table of Contents

Questions and Answers

  • S
    Susan WolfAug 8, 2025
    What to do if my Buhler Accessories has no display?
    • M
      Michael ElliottAug 8, 2025
      If your Buhler Accessories unit's display is not working, the issue could be due to a lack of power. Check the power supply to ensure it is properly connected and functioning. Alternatively, a blown fuse might be the cause. Inspect the fuse and replace it if necessary.
  • S
    sryanAug 12, 2025
    What to do if there is condensate in the gas outlet of my Buhler PKE 511?
    • L
      Laurie SmithAug 12, 2025
      If you notice condensate in the gas outlet of your Buhler Accessories, it could be due to several reasons. First, the condensate vessel might be full, so try draining the vessel. Another possibility is that the valve in the automatic condensate drain is stuck; in this case, flush it in both directions. Finally, the cooler may be overloaded, so check the limiting parameters.
  • R
    Robert ReynoldsAug 16, 2025
    Why is the status LED blinking with a high temperature on my Buhler Accessories?
    • X
      xsmithAug 17, 2025
      The status LED on your Buhler Accessories might be blinking due to high temperature for several reasons. The operational temperature might not have been reached yet; wait for a maximum of 20 minutes. The cooling capacity might be too low, so ensure free air circulation and that ventilation louvres are not obstructed. High gas flow, dew point, or gas temperature could also be the cause; check application parameters and consider installing a pre-separator. Lastly, the fan might be broken; inspect the fan and replace it if needed.
  • H
    Henry PatelAug 21, 2025
    How to troubleshoot reduced gas flow in Buhler Accessories?
    • D
      Dustin YoderAug 21, 2025
      If you're experiencing reduced gas flow in your Buhler Accessories, it could be due to a clogged gas path. Check and flush the heat exchanger. Alternatively, the condensate outlet might be clogged by ice, in which case you should send the cooler for inspection.
  • C
    careybrianAug 25, 2025
    What to do if the status LED blinks with a low temperature on my Buhler PKE 511?
    • D
      Danny PattersonAug 25, 2025
      If the status LED on your Buhler Accessories is blinking due to a low temperature, the controller may be defective. Send the cooler for inspection.
  • M
    Michael BridgesAug 29, 2025
    What to do if my Buhler PKE 511 displays error 02?
    • A
      Alyssa ChanAug 29, 2025
      If your Buhler Accessories displays error code 02, it indicates a short circuit. This means that there is a temperature sensor defect. Send the cooler for repair.
  • E
    Erin JenkinsSep 2, 2025
    What to do if my Buhler PKE 511 Accessories displays error 01?
    • L
      Lisa MendozaSep 2, 2025
      If your Buhler Accessories unit displays error code 01, it indicates a broken wire, meaning there is a temperature sensor defect. Send the cooler for repair.

Buhler PKE 511 Specifications

General IconGeneral
BrandBuhler
ModelPKE 511
CategoryAccessories
LanguageEnglish