- Check fuse and change it if
necessary
- operational temperature not yet
reached
- wait for 20 minutes maximum
- cooling capacity to low, even
though cooler is running
- make sure that air can circulate
free and that ventilation louvres
are not obstructed
- gas flow / dew point / gas tempe-
rature to high
- check application parameters,
install pre-separator
- check fan, replace if necessary
- send cooler for inspection
- Temperature sensor defect: send
cooler for repair
- Temperature sensor defect: send
cooler for repair
- valve in automatic condensate
drain is stuck
- check limiting parameters
- check / flush heat exchanger
- condensate outlet clogged by ice
- send cooler for inspection