EasyManua.ls Logo

Buhler PKE 511 - Page 37

Buhler PKE 511
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sample Gas Cooler PKE 5
The PKE Models feature a semiconductor Peltier cooling
system with an aluminum cooling block. Fitted into the block
are one or two removable high efficient heat exchanger
made of stainless steel, DURAN-glass or PVDF.
The dew point of 5 °C is regulated by an electronic controller.
The temperature (in °C or °F) of the cooling block is shown on
a programmable LED-display. The status is indicated by a
flashing display which shows too high or low temperature
and operates together with a relay in fail-safe mode.
Condensate is removed by peristaltic pumps, automatic
condensate drains or condensate vessels.
The PKE 5 is designed for moderate ambient and gas
temperatures (150 l/h @ 70°C) and an inlet dew point of
about 40 °C (appox. 5 Vol%). For higher ambient tempera-
tures up to a maximum of 50 °C order the PKE 52x.
§
§
§
§
§
§
§
§
§
Compact design
Easy installation
No maintenance required
Low noise
Model available for high ambient
temperatures
Nominal cooling capacity 90/100 kJ/h
Dew point stability 0.1 °C
Status display and status output
Cooling temperature display
DE 44 0012
Page 1/4
10/2013
Bühler Technologies GmbH
D - 40880 Ratingen, Harkortstr. 29
Tel.: + 49 (0) 2102 / 4989-0 Fax: + 49 (0) 2102 / 4989-20
Internet: www.buehler-technologies.com
e-mail: analyse@buehler-technologies.com