Maintenance
Instruction manual, Nov-17 4
Connect the unit to your power supply (1)
Set the ON/OFF switch (2) to the ON position. The Plate Heater is now ready for
use
Place the TLC/HPTLC-plate on heating plate (7)
Press the preselect button (3), at the same time turning temperature regulator (4)
in the “+” direction to increase the heating plate temperature or in the “-“
direction to lower it until the desired temperature is displayed (5)
Release the preselect button and the actual temperature of the heating plate is
displayed
The lamp (6) goes on when the plate is being heated, and goes out when the
thermostat cuts the heating
The actual temperature of the heating plate is indicated on the display (5)
3.2 Switching off the instrument
Remove the TLC/HPTLC plate from the heating plate. Set the ON/OFF switch to the
OFF position
The pilot lamp (6) goes out
As long as the unit is connected to the power supply and the heating plate
temperature is above 60 °C, the temperature is displayed
A new value can be set even with switch (2) in the OFF position
After switching back to ON, the device starts heating to the last preselected
temperature
3.3 Notes
To ensure a uniform temperature over the entire heating plate the temperature
should be set 15 minutes before use
Note that temperature and exposure time cannot simply be taken from another
type of heating such as a drying cabinet and used for the CAMAG TLC Plate
Heater III
4 Maintenance
4.1 Cleaning
To maintain proper functionality, the instrument should be cleaned regularly.
Before cleaning the heating plate disconnect the heater from the power supply
and the heating plate must be at room temperature
Mechanical cleaning: with a scraper