IX. SPECIFICATIONS
Electrical Shock Type BF Applied Part.
Environmental Conditions Transport and store 50°F to 131°F (10°C to 55°C).
Operate 41°F to 104°F (5°C to 40°C). Humidity
less than 90%. Atmospheric pressure from sea
level to 9,842 feet (3,000m)
Ingress of Water Not protected against ingress of water.
Flammability Do not use around flammable gases, liquids or
materials.
Mode of Operation Continuous.
Dimensions 6.1” x 3.5” x 1.9” (15.5 x 8.9 x 4.8cm)
Weight .4 lbs. (.18kg)
Cleaning Clean the case and lead clip wires as needed
with an alcohol moistened cloth. Do not
immerse in fluids.
Disposal Dispose of according to local, state and federal
regulations. Remove battery before disposal.
Controls Two (dose and current)
Dose Range 0 to 80 mA-min.
Maximum Voltage 80V DC
Maximum Current 4.0 mA
Current Ramp Up Automatic (0 to 4.0 mA)Built-in option for
manual override
Current Ramp Down Automatic at end of treatment; paused or turned
off by depressing knob. Built-in option for
manual override
Battery Use only 9V DC Alkaline. Ensure battery door
is in place before starting treatment. Remove
battery from unit when not in use.
Display Dose, Time Remaining, and Current(displayed
simultaneously with interactive set-up)
Pause Feature YES (w/recalculation when restarted)
Visual Indicators Low battery and open circuit
Audible Alerts
Auto Shut-Off
Low battery, open circuit, and end-of treatment
Will automatically shut off after one (1) minute, if
not in use.
7