EasyManua.ls Logo

Caputron ActivaDose II - Troubleshooting Guide; Electrode Reject Issues; Dose Limit Reached; Current Limit Reached

Caputron ActivaDose II
12 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
X. TROUBLESHOOTING
Display Shows Possible Cause Corrective Action
ELECTRODE REJECT
and“REJ” indicator lights
• Loose electrical connection
at one or both electrodes.
• One or both electrodes have
pulled away from the skin.
• Insufcient medication in
drug electrode.
• Drug electrode not properly
hydrated.
• Medication inappropriate for
iontophoresis.
• Turn CURRENT knob counter-
clockwise to “PAUSE” treat-
ment and silence beep.
• Correct problem.
• Turn CURRENT knob to reset
current. Restart treatment.
Dose “LIMIT • Maximum dose of 80.0 mA-
min has been reached.
• Refer to electrode directions
for dosage recommenda-
tions.
CAUTION: Do not exceed
recommended dosage.
Current “LIMIT • Maximum current of 4.0 mA
has been reached.
• Refer to electrode directions
for use and Section VIII-B for
recommendations on current
settings.
RESISTANCE LIMIT • Skin resistance at an
electrode site is too high
for preset current level, for
example, the plantar surface
of the foot.
• None. The unit automatically
ramps current up to preset
level, or as high as possible,
and adjusts time to delivered
desired dose.
PAUSE • Treatment has been paused
by decreasing current to 0.0
mA, or by depressing the
CURRENT knob.
• During a pause, electrodes
may be disconnected,
moved or replaced.
• Correct situation that caused
treatment to be paused.
• Turn CURRENT knob to reset
current. Restart treatment.
BAT indicator lights
during treatment.
• Battery voltage is decreasing
during treatment but treat-
ment may continue.
• After treatment is nished,
replace battery.
LOW BATTERY” and “BAT”
indicator lights when unit
is turned on.
• Battery voltage is too low for
proper operation.
• If battery voltage is very low
only BAT indicator lights.
• Unit will not allow treatment
to begin.
• Replace battery.
LOW BATT REJECT” and
during treatment “BAT” indica-
tor lights.
• Battery voltage is too low for
treatment to continue.
• Unit will not allow treatment
to be restarted.
• Unit ramps down current be-
fore preset dose is delivered.
• Replace battery.