EasyManua.ls Logo

CareFusion MicroLoop - MicroLoop Calibration Check

CareFusion MicroLoop
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Calibration Check
Connect a 3 litre syringe to the
transducer with the adapter supplied and
empty by pushing the handle fully in.
Select ‘Calibration Check from the main
menu and then select ‘check Calibration’.
Fill the syringe by pulling the handle at a
constant rate until the end stop is
reached and then immediately empty the
syringe completely. Try to maintain a
flow rate that keeps the trace within the
grey bands on the display.
Press ‘Reject’ to retry the calibration
check at the required flow rate.
Press ‘Again’ to repeat the calibration
check at a low flow rate.
Press ‘Again’ to repeat the calibration
check at a high flow rate.
When a calibration check at all three flow
rates has been completed press ‘Done’ to
view the calibration check report screen.

Related product manuals