16
The calibration error for expiration and
inspiration at each flow rate are
displayed. The calibration error should
be less that 3%. If a greater error is
shown, repeat the procedure ensuring
that the syringe is emptied and filled in a
smooth manner without jerking the
handle. If an error greater than 3% is
still shown, inspect the turbine
transducer and clean if necessary.