Screw the control valve on to the canister and connect to the
PrinterNOx with the water trap connected as shown below:
Slowly turn the control knob fully anti-clockwise. This will then
supply a flow of approximately 500ml/min. As the PrinterNOx
samples at a rate of 250ml/min then 250ml/min will be vented
through the one way valve.
Micro
Medical
Pressure
Guage
25 ppm
Nitric Oxide
in Nitrogen

OR
10 ppm
Nitrogen Dioxide
in air
One Way
Relief Val
ve
Control
Valve
Pre
-
set
Flow
Regulator
(500ml/min)
USENOOIL
0
psi
1000
4
9
Alarm
2
1
3
7
CANCEL
PAPER
6
8
CHARGING
5
0
ENTER
PrinterNOx