27
Following this, the required NO concentration in the range 1 to 100
should be entered:
EnterrequiredNO
gasconcentration
topatient:
ppm
When all the data is entered the required NO flow and the maximum
FiO
2
is calculated and displayed on the following screen:
Biasflow70L/m
Bottleconc650ppm
Patientconc20ppm
ReqNOflow2.22L/m
MaximumFiO296.9%
Presskeytocontinue
Battery operation
During patient transportation or mains power failure, the PrinterNOx
may be powered from the internal rechargeable batteries for
approximately 2 hours.