18
Instructions for Use | Carilex® VT • One
rearation to et starte
TTTIO
The battery must be fully charged prior to first use
of the VT • One suction pump. If therapy unit is in
warehouse / inventory and not used for more than
three months, the battery needs to be recharged.
VT • One is equipped with a Li-ion battery. The battery
will discharge depending on the run time of the therapy
unit and through extended periods of inactivity.
Depending on usage, the battery life cycle is 300 times
approx. Storage and usage of the battery must be
within the temperature ranges stated in the section
under “Technical Data”.
OT
lease isose of te atter in accorance it
local or facilit uielines
Collection Canister
Always make sure the canister is properly inserted and
hear "click" for proper engagement, and must remain in an
upright (carrying case) or lying (display side up) position
during use.
The VT • One suction pump is protected against the
penetration from solid / uid substances by a hydrophobic
membrane integrated together with an activated carbon
lter. If this lter fails, the VT • One suction pump must be
replaced.
1. Insert the VT • One Canister into VT • One connecting
port. Keep Canister vertical with VT • One pump (Level
marker on the left side)
2. Rotate the canister clockwise to fully engage. Please
make sure hear "click" after canister is fully engaged and
the level markers side in the same direction as display.
The VT • One system is designed for detecting when
canister is full. When the liquid absorber reaches the
canister full level, the audible and visual indicators will
be triggered and the message indicator “canister full” will
be displayed on the display panel. Do not pull the tubing
of the canister horizontally to avoid the breaking of the
suction inlet on the canister.
The VT • One collection canister is to be properly discarded
when full; it must be replaced after every patient use. The
canister should also be inspected and replaced weekly or
between patient uses or otherwise as needed.
OT
The VT • One canister is designed for single patient use
only. O OT reuse te canister to aoi te cross
infection in eteen te atients