23
Instructions for Use | Carilex® VT • One
Oeration roceure
5
Enable/ Disable the leakage alert indicator
Press and simultaneously for 3 seconds during therapy mode. Press
to select modes, until the following symbol appears on the display:
Press to select the sensitivity of leakage indicator or disable the leakage
indicator.
: Leakage alert indicator disabled
: Low sensitivity (25mmHg tolerance)
: Medium sensitivity (15mmHg tolerance)
: High sensitivity (default; with no tolerance)
After choose the appropriate sensitivity of leakage indicator, press to go
back to selected therapy mode. Adjusting the sensitivity of leakage indicator
may influence the VT• One’s function, all setting changed should be under
physician or qualified medical caregiver's order. The physician or qualified
medical caregiver should consider patient and wound's condition to determine the
appropriate sensitivity for patient.