5 Calibration
5.1. General Calibration
Pressing and holding calibration switch press [POWER] key to go to calibration mode.
User can move to other mode by using [ZERO] key in the calibration mode.
User also moves to other sub-modes for each mode by using [TARE] key.
Please simply follow below procedure to move to other mode.
(1) Calibration Mode: Pressing and holding “Calibration Switch” press [POWER] key.
(2) It displays “CAL-0” after “CAL”, and it blinks the version of scale three times.
(3) Selecting menu: press [ZERO] / press [TARE].
(4) ENTER(Setting) : [HOLD] key
MODE Function
CAL 1
Display normalized AD
CAL 2
Display Keypad information-
CAL 3
Weight Setting Mode
→
“UnLoad” [HOLD
→
]
→
“MIDD” [HOLD
→
] after loading for 1/3 weight
→
“FULL” [HOLD
→
] after loading for Full weight
→
“MIDD” [HOLD
→
] after loading for 1/3 weight “END”
CAL 4
Option Setting (Refer to the Table 1 )
CAL 5
Display filtered Raw AD
CAL 6
Function setting on each Key (Refer to the Table 2)
CAL 7
% Calibration