SPECIFICATIONS
25
Intelect® Advanced Therapy System
A series of triangular, monophasic pulses. The pulses surge to
maximum power, hold and then decrease before the pause. This
waveform is well suited for muscle strengthening.
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-80 mA
Phase Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2-5.0 ms
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-60 Hz
Surges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1/min - 20/min
Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-57 sec
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 min
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
SURGED: Monophasic Triangular
WAVEFORM SPECIFICATIONS (continued)
DANGER
Stimulus delivered by the TENS waveforms of this device, in
certain configurations, will deliver a charge of 25 microcoulombs
(µC) or greater per pulse and may be sufficient to cause
electrocution. Electrical current of this magnitude must not flow
through the thorax because it may cause a cardiac arrhythmia.

Table of Contents