What to do if my Chattanooga Intelect TranSport 2783 Medical Equipment shows Code 106?
- AArthur HolderAug 3, 2025
If your Chattanooga Medical Equipment displays Code 106, it indicates an overcurrent issue. First, check the electrodes and lead wires to ensure they are undamaged and properly connected to both the system and the electrodes. Also, verify that the electrodes are making good contact with the treatment area. If the problem persists, replace the lead wires and electrodes.