What to do if Chattanooga Intelect Genisys 2776 Intelect Genisys 2782 Intelect Genisys 2792 says no applicator is plugged in?
- MMarcus HarveyAug 12, 2025
If the Chattanooga Medical Equipment indicates that no applicator is plugged into the system, there are several potential causes. First, ensure the applicator is properly plugged into the system connector and is completely seated. If the issue persists, the applicator cable may be disconnected from the Applicator PCB; try reconnecting it. A damaged applicator cable could also be the cause, in which case, replace it with a new one. Also, the applicator might not be calibrated, so try calibrating it. Finally, a bad Ultrasound Board could be the reason, so consider replacing it.