How to troubleshoot if my Chattanooga Medical Equipment device does not work?
J
james00Aug 3, 2025
If your Chattanooga Medical Equipment device isn't working, there might be a few reasons. First, check the power supply to ensure it's properly connected. If the power supply is fine, the mains fuse might be defective, in which case you should replace the fuses. Lastly, a defective mains plug could also be the culprit, so try replacing the mains cable.
E
Evan TurnerAug 13, 2025
What to do if there is no F-SW power output in my Chattanooga Medical Equipment?
A
Aaron BoothAug 13, 2025
If you are experiencing a situation where there is no F-SW power output from your Chattanooga Medical Equipment, it could be due to a defective F-SW handpiece or a malfunction in the control device. Try replacing the handpiece first. Also, if the handpiece hasn't been recognized, ensure the black screw is properly tightened.
J
John VasquezAug 22, 2025
Why does the shock triggering noise change after several shocks in my Chattanooga Medical Equipment?
L
Laura NixonAug 22, 2025
If the shock triggering noise changes after several shocks with your Chattanooga Medical Equipment, it's likely due to air in the handpiece. To resolve this, hold the handpiece vertically with the coupling diaphragm downwards to allow the air to be sucked out.