TechnicalSpecications
EMC guidelines and manufacturer’s declaration
Guidelines and manufacturer's declaration – emitted electromagnetic interference
The Chattanooga Intelect F-SW model is intended for operation in the electromagnetic environment
specied below. The customer or the user of the Chattanooga Intelect F-SW should ensure that it is used
in such an environment.
Interference emission
measurements
Compliance Electromagnetic environment – guidelines
HF emissions acc.
to CISPR 11
Group 1
The Chattanooga Intelect F-SW uses HF energy only for
its internal functioning. Therefore, its HF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment. According to EN IEC 60601-
2-36:1997 Section 36 this does not apply during the
generation and release of the pressure pulse
HF emissions acc.
to CISPR 11
Class B
The Chattanooga Intelect F-SW is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
according to IEC 61000-
3-2
Class A
Voltage uctuations
/ icker emissions
according to IEC 61000-
3-3
Complies