Cleaning, Maintenance and Overhaul
6.5 Fuse replacement
The mains fuse holder is located on the rear panel of the Chattanooga Intelect F-SW.
• Push the clip of the mains fuse holder to the right and take the holder off the
housing.
1
Fig. 6-62 Mains fuse holder
• Pull the old fuses out of the mains fuse holder.
1
Fig. 6-63 Fuse replacement
• Replace the fuses (T 5 AH / 250 VAC).
• Push the mains fuse holder back into the opening until it engages.
6.6 Maintenance and safety checks
Preventive maintenance is not necessarily required. However, regular maintenance
may help to identify possible defects at an early stage and thus increase the safety and
service life of the instrument.
Maintenance services can be ordered from our regional representatives in your area.
We recommend that functional and safety checks be performed at least once a year.
National accident prevention regulations and test and inspection intervals prescribed
for medical devices must, of course, be observed.
The following checks should be performed to ensure that the Chattanooga Intelect
F-SW operates safely.
1 Earth leakage current test according to national regulations.
2 Earth impedance test (with mains cable, incl. applicator housing) according to nati-
onal regulations.
3 Checks essential performance
NOTE
For further details on content and performance of the safety checks please con-
tact your local dealer.