5.15 Functional checks
Perform the following functional checks after the system has been installed:
Check the control unit and handpieces for damage.
Start the Chattanooga Intelect F-SW (see chapter 5.14 start-up).
Set the energy level in F-SW mode to 0.2 mJ/mm
2
.
Optional: Set the energy level in C-ACTOR mode to 0.69 mJ/mm
2
.
Reset the actual number of shocks on the parameter display of the control panel
(see chapter 5.12 resetting the treatMent shOck cOunter).
Release shocks with a shock frequency of 4 Hz.
Release shocks by means of the foot switch, if used.
Check that the triggered shocks are correctly counted on the treatment shock
counter.
NOTE
If necessary, the functional capability of the F-SW handpiece can be checked with
the aid of special Colour sensitive pressure sensors sensors.
5.16 Standard settings
Before each treatment, make sure that the number of shocks and the actual en-
ergy value are set to zero (see chapter 5.12 resetting the treatMent shOck cOunter).
NOTE
Set the nominal value counter to the required value. The "-" symbol appears if
zero is selected. The instrument then operates without a nominal value specica-
tion.
Start the F-SW treatment at an energy level of 0.1 mJ/mm
2
and a frequency of 6
Hz.
Start the C-ACTOR treatment at an energy level of 0.03 mJ/mm
2
and a frequency
of 6 Hz.