Cleaning, Maintenance and Overhaul
6.4.2 Filling the water circuit
Make sure that the instrument is standing on a smooth, horizontal surface.
Rinse out the water bag.
Use only deionised water (in compliance with VDE 0510, e.g. water for batteries or
clothing irons) to rinse or ll the water bag.
Fill the water bag to the brim.
Fig. 6-57 Filling the water bag
ATTENTION
Do not use water that has been distilled more than once!
After the water bag has been lled, there should be as few bubbles as possible
in the connection tube. Press the closing valve inwards to let the air escape (see
picture below) until the hose is fully lled with water.
Fig. 6-58 Venting the connection tube
Place the F-SW handpiece into the F-SW handpiece holder so that any air bubbles
that form will be immediately sucked up by the bubble trap.
Activate “Fill water circuit” operating mode in the “Setup” menu
Connect the water bag to the water tube connection on the rear of the instrument
as soon as the message appears.