Status messages and trouble-shooting
7 Status messages and trouble-shooting
7.1 Status messages
CAUTION!
Malfunction of the device or its components
Various injuries are possible!
Immediately comply with all status and error messages which
appear during the treatment.
Speciednumberofshock
waves reached
Acknowledge message,
further treatment is possible.
Shock wave safety stop
Acknowledge message,
further treatment is possible.
F-SW: load test
unsuccessful
Restart the instrument and repeat
the test.
Do not continue to use the
instrument if the fault continues.
Notify your Service centre
F-SW: charging timeout
Acknowledge message.
Inform your Service centre if the fault
continues.
F-SW: water temperature
too high
Acknowledge message,
further treatment is possible once
the water temperature has returned
to permitted values.
F-SW: water temperature
too low
Acknowledge message,
further treatment is possible once
the water temperature has returned
to permitted values.
F-SW: water level too low
Fill the water circuit (see chapter
6.4.2 filling the water circuit)
F-SW: water circuit fault
Pump defect
Treatment is not possible.
Bleed the water circuit (see chapter
6.4.3 bleeding the water circuit).
Inform your Service centre if the fault
continues.
F-SW: water pump current
too low
Acknowledge message.
Inform your Service centre if the fault
continues.
F-SW: pump temperature
too high
Acknowledge message,
further treatment is possible once
the pump temperature has returned
to permitted values.