Status messages and trouble-shooting
F-SW: therapy head
overtemperature
Acknowledge message,
further treatment is possible once
the therapy head temperature has
returned to permitted values.
F-SW: water temperature
sensor failure
Restart the instrument.
Inform your Service centre if the fault
continues.
Shock wave limit for
current handpiece reached
Shock wave limit for current
handpiece reached. Replace the
handpiece.
F-SW: charging unit not
ready
Acknowledge message.
Call your Service centre if the fault
continues after a reset.
USB stick was not
recognised
Remove the USB stick, then switch
off and restart the instrument.
Reinsert the USB stick.
Check that there is software on the
USB stick.
If the fault persists, check that the
USB stick supports the USB V1.1
protocol. If it does not, replace the
USB stick.
Water amount
insufcient,pleasecheck
supply
Fill the water bag and check whether
water is owing into the water
circuit.
If the fault persists, inform your
Service centre.