13 610 02 1018
61
Cleaning, Maintenance and Overhaul
.
Fig. 6-54 Draining the water circuit I
The “Please wait” message and a progress indicator appear on the display.
Allow the remainder of the water to drain out of the handpiece by holding the
F-SW handpiece vertically above the instrument as soon as you are prompted to do
so. Make sure that the coupling diaphragm of the handpiece is pointing upwards.
Fig. 6-55 Draining the water circuit II
The “Please wait” message and a progress indicator appear on the display.
Wait until the instrument is ready. The display shows when the water circuit is
empty.
Fig. 6-56 Draining the water circuit III
Open the lock on the tube connection and pull the tube out of the tube connec-
tor.
Remove the full water bag and dispose of the contents.