EasyManuals Logo

Checkme O2 Oxiband User Manual

Checkme O2 Oxiband
28 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
19
Harmonic emissions
IEC61000-3-2
Class A
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Voltage Fluctuations / Flicker
Emissions IEC 61000-3-3
Complies
Guidance and Declaration - Electromagnetic Immunity
The Health Monitor is intended for use in the electromagnetic environment specified below. The
customer or the user of the Health Monitor should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance level
Electromagnetic
environment -
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic material,
the relative
humidity should
be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Mains power
quality should be
that of a typical
commercial or
hospital
Surge
± 1 kV line(s) to
± 1 kV line(s) to

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Checkme O2 Oxiband and is the answer not in the manual?

Checkme O2 Oxiband Specifications

General IconGeneral
BrandCheckme O2
ModelOxiband
CategoryMonitor
LanguageEnglish