30
EN
Schülke &
Mayr
Mikrozid
Liquid
1 to 2mins
Based on the following material com-
patibilities, you can assess which other
cleaning agents and disinfectants are
suitable.
The anodised aluminium components of
the SOL Nova are resistant to:
Weak acids (e.g. acetic acid ≤10%, citric
acid ≤10%)
Aliphatic fluorocarbons (e.g. pentane,
hexane)
Ethanol
Most inorganic salts and their aqueous
solutions (e.g. sodium chloride, calcium
chloride, magnesium sulphate)
The anodised aluminium components of
the SOL Nova are not resistant to:
Strong acids (e.g. hydrochloric acid,
sulphuric acid, nitric acid)
Oxidising acids (e.g. peracetic acid)
Lyes (e.g. caustic soda, ammonia and
all substances with a pH value >7)
Aromatic/halogenised fluorocarbons
(e.g. phenol, chloroform) as well as
acetone and benzene
7.2 Disinfecting surfaces
Before and after each usage:
! Disinfect all surfaces of the SOL Nova
and its accessories as described in
chapter “Disinfecting equipment” on
page25.
7.3 Cleaning surfaces
In the event of external contamination:
! Wipe the SOL Nova and the protective
hose with a soft cloth dipped in a
cleaning agent or a disinfectant.
! Clean inaccessible surfaces, e.g.the
thread of the drive unit, with a soft-
bristled nylon brush.
7.4 Cleaning handle in an ultrasonic
bath
If the handle is severely soiled:
! Use the cleaning agent Tickomed 1
or Tickopur R33 from manufacturer
Dr. H. Stamm GmbH in accordance with
the manufacturer's instructions.
! Make sure that a pH value of 7 is not
exceeded.
! Conduct the cleaning operation for
10minutes at a frequency of 35kHz.
! Dry the handle fully after cleaning,
before you use it again.
7.5 Sterilising handle in an
autoclave
Whenever the handle is contaminated:
! Perform all operations described in
chapter 'Disinfecting equipment' (see
Page25).
! Autoclave the handle for 20minutes at
a temperature of 121°C and a pressure
of 2bar.
! Dry the handle fully after cleaning,
before you use it again.
7.6 Cleaning or replacing O-rings
There are two O-rings that prevent the
handle from turning too smoothly. If it
is too easy to turn the handle so that the
needle protrusion alters accidentally:
! Degrease the O-ring with a soft
cloth dipped in a cleaning agent or
adisinfectant.
! If the problem persists, replace
theO-rings with new ones and lubricate
them with silicone grease if necessary.
Cheyenne SOL Nova