EasyManua.ls Logo

Cholestech LDX - 8 Glossary of Terms

Cholestech LDX
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38 GLOSSARY OF TERMS
Cholestech LDX System Procedure Manual
< TABLE OF CONTENTS
Acceptable Control Range – range of results that indicate adequate performance when
analyzing a control sample. The range is shown in the control’s product insert.
Accuracy – correctness; freedom from error. The accuracy of results can be measured by
comparing them with those from another laboratory (this is “
relative accuracy
”).
Additive – chemical added to a blood collection tube, usually to prevent the blood from
clotting (anticoagulant).
Aerosol – fine mist that solid or liquid particles are dispersed in.
Agglutination – clumping together of antigen-bearing cells, bacteria, or particles in the
presence of specific antibodies. Also called “
clumping
.”
Aliquot – small portion of a measured volume of a substance taken as a sample
representing the whole.
Analysis – laboratory procedure that enables measurement of the amount of an analyte in
a specimen.
Analyte – substance or constituent being measured (
e.g.,
cholesterol, triglycerides, glucose).
Antibody – substance formed in the body in response to a foreign substance (an
antigen
)
and that interacts only with that substance.
Anticoagulant – chemical used to prevent blood from clotting.
Antigen – any substance that, injected into an organism, causes the development of
antibodies.
Antiserum – serum that contains antibodies.
Aseptic – free from infection or septic material; sterile.
Assay – measurement of the amount of an analyte in a specimen; a test.
Autoclave – instrument that sterilizes material by subjecting it to steam under pressure.
Bias (inaccuracy) – measure of the departure from accuracy. A numerical difference
between the mean of a set of replicate measurements and the true value of the sample.
Calibrated – (of a measuring device,
e.g.,
a pipette) graduated into appropriate units.
Calibration – taking readings from an instrument or other measuring device and relating
them to known concentrations of an analyte or true value.
Calibrator – material, solution, or freeze-dried preparation used in calibration. The
concentration of the analytes in a calibrator is known to be within a particular range.
Calibrators may be a primary or a secondary standard.
Capillary – any one of the small vessels that form a network throughout the body for the
interchange of substances between the blood and tissue fluid.
Glossary of Terms
I