Digital Power Meter 66203/66204 User’s Manual
5-38
Query syntax: none
Return parameters: none
Example: none
SHOW[:DISPlay]:ITEM
Type: System.
Description: This command is used to select which item of measure will be
displayed.
Setting syntax: SHOW:ITEM<space><arg1>,<arg2>,<arg3>,<arg4><PMT>
Setting parameters: <arg1 ~ 4> denote four display areas, and there are in <CRD> type,
arg1: V, I, W, IS, VPK+, PF, F
arg2: V, I, W, IS, VPK-, EFF, CFI
arg3: V, I, W, IS, E, THDV, THDI
arg4: V, I, W, PF, VA, VAR, (Ew, Epf, Eva, Evar only for non-1P2W
mode)
Query syntax: none
Return parameters: none
Example: none
CALCULATE Sub-system
[CALCulate:]COMParator
Type: All channels.
Description: Users can configure upper and lower boundary of measured items.
662xx will check measured items according to the boundaries. Once
any item exceeds the boundaries, the item will be recorded. The
command is used to turn on/off the comparison function.
Setting syntax: COMParator<space><CRD><PMT>
Setting parameters: <CRD>, ON | OFF
Query syntax: COMParator?<PMT>
Return parameters: <CRD>, ON | OFF
Header on: :COMPARATOR<space><CRD><RMT>
Header off: <CRD><RMT>
Example: none
[CALCulate:]COMParator:TIME
Type: All channels.
Description: Programming “COMP:TIME” decides the dwelling time of
comparison function. The unit of Setting Parameter is second.
Setting syntax: COMParator:TIME<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 9999
Query syntax: COMParator:TIME?[<space><MAX | MIN>]<PMT>
Return parameters: <NR1>, 0 ~ 9999
Header on: :COMPARATOR:TIME<space><NR1><RMT>
Header off: <NR1><RMT>
Example: none
[CALCulate:]COMParator:RESult?
Type: All channels.