ELECTOROMAGNETIC COMPATIBILITY INFORMATION
WARNING
Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the [CH-453], including cables specied.
Use of this equipment adjacent to or stacked with other equipment should be avoided.
Use of accessories and options other than those specied (other than CITIZEN
original parts) could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment.
Guidance and manufacturer’s declaration - electromagnetic emissions
The [CH-453] is intended for use in the electromagnetic environment specied below. The
customer or the user of the [CH-453] should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The [CH-453] uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The [CH-453] is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage uctuations/ icker
emissions IEC 61000-3-3
N/A
Guidance and manufacturer’s declaration - electromagnetic immunity
The [CH-453] is intended for use in the electromagnetic environment specied below. The
customer or the user of the [CH-453] should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If oors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply
lines
±1 kV for
input/output lines
N/A N/A
Surge
IEC 61000-4-5
±1 kV
line to line
±2 kV
line to earth
N/A N/A
Voltage dips,
short
interruptions
and voltage
variations on
power supply
IEC 61000-4-11
0% U
T
0.5 cycle
N/A N/A
0% U
T
1 cycle
70% U
T
25/30 cycle
0% U
T
250/300 cycle
Power frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
30 A/m 30 A/m
Power frequency magnetic elds should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: U
T
is the A.C. mains voltage prior to application of the test level.
CH-4531901.indb26 2019/05/3113:45