EasyManua.ls Logo

CleanWater Tech CD1500P - Ozone Injection and Contacting Issues

Default Icon
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting (continued)
22
OZONE GENERATOR
continued
PROBLEM/SYMPTOM POSSIBLE CAUSE SOLUTION
Unit flooded with water - Defective check valve - Replace check valve(s)
Ozone smell detected in or - Assess damage, repair as required near
ozone generator - Insufficient vacuum at venturi - Adjust injector See “Start- Up &
Calibration - Steps 1-4”
- Loose internal fittings - Check all fittings, tighten as needed
- Defective O-ring seals in reaction chamber(s) - Check & replace as required
- Defective dielectrics - Check & replace as required
OZONE INJECTION/CONTACTING
PROBLEM/SYMPTOM POSSIBLE CAUSE SOLUTION
Water backflow past injector - Defective check valve - Replace check valve
check valve
Water bubbling in vacuum break - No vacuum - See “Start-Up & Calibration”
- Debris on seat of vacuum break flapper valve - Clean seat of flapper. SeeMaintenance
Procedures - Annual
Low vacuum - Hydraulics/Pneumatics out of adjustment - SeeStart-Up & Calibration - Steps 1-4”
- Defective check valve - Replace check valve
- Hydraulic back pressure - Back wash filter (if so equipped)
- Check for obstruction in venturi
- Defective solenoid valve - Rebuild or replace as required
- No water in vacuum break - Fill vacuum break with water See
Start-Up and Calibration Vacuum
Break
- Booster pump not functioning properly - Check booster pump (contact dealer)
High Vacuum - Hydraulics/Pneumatics out of adjustment - See “Start-Up & Calibration Steps 1-4”
- Change in hydraulics excessive water flow - See “Start-Up & Calibration Steps 1-4”
through ozone injector
Ozone smell detected around - No vacuum - See “Start-Up & Calibration Steps 1-4”
vacuum break or ozone injector - Loose fittings - Tighten all fittings
- Broken fittings - Replace fittings