12. Error messages
Error No. Description Remedy
00 Electronics overheating–safety shutdown Switch off the equipment, allow to cool for at least
10 minutes, then restart
01 Electronics overloaded Switch off the equipment and then restart
07 Foot control errorinitialising Switch off the equipment, restart, do not actuate
foot control when switching on
09 Foot control error Switch off the equipment, check connection of the
foot control, then restart
19 Run limiting control Switch off the equipment and restart
99 System failure Switch off the equipment and restart
Motor temperature too high Switch off the equipment, allow motor to cool for
at least 10 minutes, then restart
Motor connectionsafety shutdown Switch off the equipment, check motor connection,
then restart