17. Technical data
Motor and foot control conform to Class AP according
to IEC 60601-1/UL 60601-1 in danger zone M.
The foot control is watertight according
to PX8, 1m depth of immersion, 1 hour
(water-tight according to IEC 60529).
Classification according to § 5 of the General Specifications for
the Safety of Medical Electrical Equipment according to IEC
60601-1/UL 60601-1:
Equipment of Protective Class II (type of protection
against electric shock).
Type BF appliance (not suitable for intracardiac
application).
91051 / 91052 91053
Supply voltage: 220–240 V 100–130 V
Rated current: 0.1–0.8 A 0.2–1.6 A
Permitted voltage fluctuation: ± 10 % ± 10 %
Frequency: 50–60 Hz 50–60 Hz
Max. mechanical output power: 70 W 70 W
Max. torque on the motor: 5.5 Ncm 5.5 Ncm
Max. coolant in ml (with hose ø 6/4): approx. 100 ml/min approx. 100 ml/min
Fuses: 2x 250 V–1.25A slow-blow 2x 250 V–1.25A slow-blow
Operating mode: S3 (4min/10min) S3 (4min/10min)
Dimensions in mm (WxDxH): 252 x 254 x 90 252 x 254 x 90
Set weight in kg: 4.10 4.10