13. Disinfection, cleaning, sterilisation
Motor with cable
Do not twist and kink the motor cable! Do not coil
it too tightly! This instruction is not necessary when
a sterilisation cassette is utilised (optional accessory).
Before disinfection and cleaning push the motor
protective cover onto the motor attachment!
Pre-disinfection
– If heavily soiled: Clean first with disinfectant cloths.
Only use disinfectants that have no protein-fixing effects.
Manual cleaning
Rinse under demineralised water (< 38 °C) with the aid of
a brush (brush is not recommended in U.K.)
Then remove any liquid residues (absorbent cloth, blow dry
with compressed air).
Disinfection
Disinfect with surface disinfectants. Wiping disinfection
is recommended.
Use only surface disinfectants which do not contain chlorine
and which are certified by officially recognised institutes.
For USA: Use EPA registered surface disinfectants.
Note the manufacturer’s specifications for the use of the
surface disinfectants.