14. Osscora contra-angle handpiece
Manual disinfection
Disinfect with surface disinfectants.
u Wiping disinfection is recommended.
Use only surface disinfectants which do not contain chlorine
and which are certified by officially recognised institutes.
For USA: Use EPA registered surface disinfectants.
Note the manufacturer’s specifications for the use of the
surface disinfectants.
After manual cleaning, disinfection and lubrication
you must carry out a final sterilisation (wrapped) in
the class B or S steam steriliser (according to EN
13060 or EN 285).