14. Osscora contra-angle handpiece
Sterilisation and storage
Cochlear recommends sterilisation according to EN
13060 or EN 285, class B.
Note the instructions of the steriliser manufacturer.
Clean, disinfect, reassemble and lubricate before sterilising.
Wrap the contra-angle handpiece and the accessory in sterile
goods packing according to EN 868-5.
Make sure, that you only remove dry sterile goods.
Store sterile goods dust-free and dry.
Approved sterilisation procedures
Follow your country-specific directives, standards and
guidelines.
Steam sterilisation with sterilisers in accordance with
EN 13060 or EN 285, class B.
Sterilisation holding time: Minimum of 3 minutes at
134 °C (273.2 °F)
or
Steam sterilisation with sterilisers in accordance
with EN 13060 or EN 285, class S.
The steriliser manufacturer must give their express approval
for the sterilisation of contra-angle handpieces.
Sterilisation holding time: Minimum of 3 minutes at
134 °C (273.2 °F).