32
Chapter 4 Heater Control
Advanced Settings
Skipping soaking steps
When initiating a heat cycle, the user has the option of skipping Soak and Soak 2 and
initiating the heating cycle directly. The following settings will allow the heater to enter
PID mode to achieve the desired temperature (in this instance a temperature of 125°C):
Clean Settings
Max Power [%] 100
Proportional 15
Integral 5
Derivative 0
Rise to Soak Time [s] .1
Soak Time [s] 0
Soak Power [%] 0
Rise to Soak 2 Time [s] .1
Soak 2 Time [s] 0
Soak 2 Power [%] 0
Stabilization Time [s] 0
Heater Set Point [C] 125
Clean Time [s] 60
Cool Time [s] 30
Bypass settings